Evaluation of Tensile Strength of Sutures Used in Dentistry

dc.contributor.authorErcal, Pinar
dc.contributor.authorTaysi, Aysegul Erten
dc.contributor.authorTaysı, Nurı Mert
dc.contributor.authorSismanoglu, Soner
dc.date.accessioned2026-02-06T17:48:03Z
dc.date.issued2023
dc.departmentDoğu Akdeniz Üniversitesi
dc.description.abstractAim: The aim of this study was to compare the mechanical properties of sutures used in dentistry according to different materials in vitro. Material and Methods: Eight 3-0 different absorbable and non-absorbable suture materials with 3-0 gauge (Polytetraflouroethylene, polypropylene, polyester, polyglactin 910, polyglycolic acid, poliglecaprone 25, polydioxanone, and silk) were compared in terms of mechanical strength. All sutures were tied with a simple suture technique. Each material contained 10 samples per group, with a total sample size of 80 specimens. Failure load was measured in N while elongation was measured in µm using a microtensile testing device. A one-way analysis of variance (ANOVA) was used to analyze the difference in failure loads and elongation values. Results: Polydioxanone was significantly more resistant to tensile forces among all tested materials (p
dc.identifier.doi10.33631/sabd.1146151
dc.identifier.endpage193
dc.identifier.issn2792-0542
dc.identifier.issue2
dc.identifier.startpage189
dc.identifier.trdizinid1171170
dc.identifier.urihttps://doi.org/10.33631/sabd.1146151
dc.identifier.urihttps://search.trdizin.gov.tr/tr/yayin/detay/1171170
dc.identifier.urihttps://hdl.handle.net/11129/6687
dc.identifier.volume13
dc.indekslendigikaynakTR-Dizin
dc.language.isoen
dc.relation.ispartofSağlık bilimlerinde değer (Online)
dc.relation.publicationcategoryMakale - Ulusal Hakemli Dergi - Kurum Öğretim Elemanı
dc.rightsinfo:eu-repo/semantics/openAccess
dc.snmzKA_TR-Dizin_20260204
dc.subjectTensile strength
dc.subjectsutures
dc.subjectmechanical testing
dc.titleEvaluation of Tensile Strength of Sutures Used in Dentistry
dc.typeArticle

Files